4 preparation (setting parameters) – A&D Digital Platform Scales HW-10KGV User Manual

Page 57

Advertising
background image

HW-G Instruction Manual Page 55

HW-G OZIM v1a

16.4 Preparation (Setting Parameters)

Step 1

Turn off the display.
Press the

ON/OFF

switch while the

ZERO

switch is

pressed and held. The function table is displayed.

Step 2

Select a parameter of the “baud rate (

f4

)” using

the

ENTER

and

^

switches.

Step 3

Select “command mode (

f5 1

)” of the “output”

using the

ENTER

and

^

switches.

Step 4

Select “format 1 (

f15 0

)” of the “format” using

the

ENTER

and

^

switches.

Step 5

Press the

ENTER

switch to store the new

parameters.
Return to the normal weighing mode using the

ENTER

switch.

Step 6

Receiving a command, the scale responds.