A&D Digital Platform Scales HW-10KGV User Manual

Page 59

Advertising
background image

HW-G Instruction Manual Page 57

HW-G OZIM v1a

17.2 RS-422 / RS-485 / Relay output (OP-04)

Replacing RS-232C interface with this option, the RS-422/ RS-485 interface can connect
up to 16 scales and control them from a computer.

The RS-422/ RS-485 interface has the following two modes.

Stream mode

Outputs data countinuouslly.

Command mode

Controls the scale using commands from a computer.

It is necessary to set the parameters of the “Baud rate (

f4

)”, “

Output mode (

f5

)”, “ Format (

f15

)” in the function table in advance.

Transmission systemEIA RS-422/ RS-485
Transmission

Asynchronous, bi-directional, half-duplex

Data format

Baud rate: 2400, 4800, 9600 bps
Data:

7 bits

Parity:

1 bit, Even

Start bit

1 bit

Stop bit

1 bit

Code

ASCII

Terminator

CR LF (CR: 0D[H], LF: 0A[H])

Address

01 ~ 99

Address parameter (

f10

) of function table.

Relay

Solid-state-relay
Maximum voltage

DC50V

Maximum current

DC100mA

Maximum resistance 8

Mating connector

TM:BLA9 (of accessory pack)

Pin connections

Pin No. Signal name

Direction

Description

1

SDA

Output

RS-422 transmission A terminal

2

SDB

Output

RS-422 transmission B terminal

3

RDA

Input

RS-422 receive A terminal

Input / Output

RS-485 transmission/ receive A
terminal

4

RDB

Input

RS-422 receive B terminal

Input / Output

RS-485 transmission/ receive B
terminal

5

TRM -

120

terminator

6

HI

Output

Relay output of HI

7

OK

Output

Relay output of OK

8

LO

Output

Relay output of LO

9

COM -

Relay common terminal

Advertising