HW-G Instruction Manual Page 42

HW-G OZIM v1a

14.2.3 Calibration of the Zero Point

Step 8 Press and hold the

CAL

switch to

enter the calibration mode after
displaying normal
weighing for 30 minutes.
Then the

Cal 0

is displayed.

Step 9 With nothing on the pan, press the

ENTER

switch while the stable

mark is displayed. The scale
stores the current condition as the
zero point. .

Step 10

The scale displays

5pn1

for

several seconds. If finishing
the calibration mode at this
stage (Zero calibration only),
press CAL to proceed to step 13.

14.2.4 Span Calibration

Step 11 Set the value of the calibration

mass using the following switches.
(The initial value depends upon the
particular model.)

^

switch

Selecting the value of

a digit.

<

switch

Selecting a digit.

Step 12 Place the mass’ selected at

step 11’ onto the pan. When the
stable mark is displayed press
the

ENTER

switch.

The scale then calculates the
span and stores the value.

Step 13 The scale displays

end

at the

finish.
Remove the mass from the pan.

Step 14

Press the

CAL

switch to return

to the normal weighing mode.