1 data format, Format 1 – A&D Digital Platform Scales HW-10KGV User Manual

Page 51

Advertising
background image

HW-G Instruction Manual Page 49

HW-G OZIM v1a

16.1 Data Format

Bit Format

Format 1

There are four headers for the type of data and weighing condition.

Stable weighing data

S T

Unstable weighing data

U S

Stable counting data

Q T

Out of range (over)

O L

The weighing data consists of 9 characters including decimal point and polarity.

The polarity is always output.

There are three units. The unit is 3 characters.

In case of “out of range”, numbers become all 9

s.

The terminator is always output as CR LF.

Data in the unit of kg

Counting mode

Percentage mode

Out of range

Definition of symbols

CR (carriage return)

C

R

LF (line feed)

L

F

0D[H]

0A[H]

Space
20[H]

Advertising