CiDRA GVF-100 Gas Volume Fraction Monitoring System User Manual

Page 103

Advertising
background image

Copyright © 2006 CiDRA Corporation

Page 13-2

20639-01 Rev 03

10. Install a sensor band shorting plug on the sensor band cable. This

should be stored in the transmitter enclosure. Note: Sensor bands
with an “R” in the Part Number suffix (e.g. Part #: 20686-26-R) do
not require a shorting plug.

WARNING

Potential electrical discharge. Always install a sensor band
shorting plug (see item 10 above) on the sensor band connector
whenever the sensor band connector is not installed in a mating
connector.

10. Install spring clamps on both sides of the cover flanges to keep it

from sliding or separating while removing the cover bolts.

11. Remove the boot gasket band if removing a stainless steel cover.

12. Remove the sensor cover bolts, nuts, and washers, and then cover

from pipe.Replace sensor cable access cover.

14. Use a hex driver or Allen wrench to unscrew the nine sensor

screws that hold the two rails of the sensor together. Unscrew
each screw a couple of turns until they are loosened. While the
screw threads are still engaged in the attachment rails a few turns,
spread the sensor rails apart and then continue to loosen the
screws until they are no longer engaged in the threads. (By doing
this, the o-ring that keeps the screws from coming out of the rails
will stay in place.) Use care to keep the fastener assembly intact.
If the spacers or Bellville washers do come off the screw refer to
the sensor band screw figure in the Sensor Installation section of
the manual for proper stack-up of components.

15. Remove the sensor band assembly from the pipe.

16. Slide the screw and washer assembly o-ring down the screw to

retain the screws in their holes.

17. The original packaging as described in Section 6.1 should be used

when storing/transporting the system. If not available, lay the
sensor band in the bottom of the cover and move it to an area
where it will not be damaged.

18. Replace sensor band screws prior to re-installation of sensor band.

19. Inspect and replace fiberglass cover gaskets as required.

20. Inspect and replace stainless steel cover flange gaskets as

required.