CiDRA GVF-100 Gas Volume Fraction Monitoring System User Manual

Page 67

Advertising
background image

Copyright © 2006 CiDRA Corporation

Page 9-16

20639-01 Rev 03

Table 13

Transmitter Menu Tree Software Release 03.03.XX

Level 1

Level 2

Level 3

Range

Description

Sensor
Serial #

0000000

Serial number of sensor
band

ID/Wall

ID: 1- 100 in (25.4-
2540mm)

Wall 0 – 100 in (0 to
2540 mm)

Pipe inner diameter and
wall thickness

Size/Sched

2 to 36" size;
schedule

Pipe size & schedule

Pipe Size

OD/Wall

ID: 1- 300 in (25.4-
7260mm)

Wall 0 – 100 in (0 to
2540 mm)

Pipe outer diameter and
wall thickness

Pipe Material

SST, CS, PVC, Custom

Enters the pipe
modulus, kilo-Pascal

Spec Gravity

0 - 999999

Enters the specific
gravity; default water at
25 °C and 14.7 psia

Fluid
Properties

SOS (ft/s)

0 - 999999

Sound speed in media
of interest; ft/sec;
default water at 25 °C
and 14.7 psia

Pressure

+/-

0-999999

Process pressure; PSIg,
Barg, kPag

Temperature

-999 to +999C
-1766 to 1830F

Process temperature;
°C or F

Pressure Sel

Fixed, Sensor #1 or #2

Allows for selecting a
fixed pressure input or
using the inputs of
sensor #1 or #2

Temperature
Sel

Fixed, Sensor #1 or #2

Allows for selecting a
fixed temperature input
or using the inputs of
sensor #1 or #2

Altitude

-50,000 to +50,000

Process pipe altitude
above / below sea level;
feet or meter

Basic Config

Set
Date/Time

MM/DD/YY
HH:MM:SS

Current date/time;
adjust for daylight
savings as required

Advertising