CiDRA GVF-100 Gas Volume Fraction Monitoring System User Manual

Page 61

Advertising
background image

Copyright © 2006 CiDRA Corporation

Page 9-10

20639-01 Rev 03

Status Messages - ‘S’:

A number of status messages can be displayed in the 20 character
status message field. They are as follows:

Status Message

Description

GF INITIALIZE MODE

GVF Initialize mode

INVALID SOS DATA

Invalid SOS data

BELOW MIN SS QUALITY

SOS quality below configured threshold

Table 9

Status Line Messages

Quality Messages ‘Q’:

The Quality Field is a diagnostic field that can be used to view certain
quality values in the transmitter. The factory default is none. If more
than one quality is selected, the transmitter will cycle through each.
The options for display are speed of sound, pressure and temperature
(if used), band temperature, and a 3-level quality metric. The 3-level
fields represent a Red/Yellow/Green setting for the quality of the
output data. The yellow quality is the bounded value where data can
be considered valid, but is not fully reliable. Red indicates unusable,
and green indicates a good measurement.

Quality Message

Description

SOS ‘-1 to +1’

Quality metric of the measurement is between –1 and +1;
+1 is the most robust measurement system can make

RED

Quality metric is below the minimum quality factor entered
into the transmitter or the meter is in startup mode

YEL

Quality metric is usable but not highly reliable

GRN

Quality metric is highly reliable

B

Band temperature (°C)

T

Process temperature (either from remote sensor or
programmed into system, user selectable °C or F) (if used)

P

Process pressure (either from remote sensor or
programmed into system, user selectable units) (if used)

Table 10

Quality Message Definitions

Mode Messages – ‘M’:

The four character mode field is used to display the operating mode of
the system. Their definitions are as follows:

Mode Message

Description

‘IDL’

Idle / Stop Mode

‘RAW’

Transferring raw data

‘SNG’

Performing a single measurement

‘GVF/SOS’

Running in Gas Volume Fraction / Speed Of Sound mode

Table 11

Mode Message Definitions