2 boot gasket fender assembly installation, 3 sensor assembly cable connection, 3 sensor assembly cable connection -18 – CiDRA GVF-100 Gas Volume Fraction Monitoring System User Manual

Page 35

Advertising
background image

Copyright © 2006 CiDRA Corporation

Page 7-18

20639-01 Rev 03

7.6.2.2

Boot Gasket Fender Assembly Installation

The stainless steel cover uses four boot gasket fender assemblies
(two per end) to protect the boot gaskets. The fender should be
installed such that the retaining clasp is next to the cover handles.
The completed cover assembly is shown in Figure 17.

Figure 25

Boot Gasket Fender Assembly

7.6.3

Sensor Assembly Cable Connection

Remove the tape (if it was used) that was temporarily installed to
retain the sensor connector under the cover access panel. Remove
the factory installed shorting plug from the sensor cable connector.
Place the shorting plug in the holder in the transmitter housing. (It will
be used if sensor removal from the pipe is necessary.) Install the
sensor assembly cable connector into the mating receptacle located
inside the access cover plate on the sensor upper cover as shown
below. The connector is installed such that the connector key faces
away from the pipe. Push the connector into the mating receptacle.
The locks on the receptacle will engage to keep the plug from pulling
out. Ensure the connector is fully engaged. Note: Use care to ensure
the sensor cable connector goes on squarely to minimize risk of
damaging the contacts.

Figure 26

Sensor Assembly Cable Installation

Connector locks

Sensor cable
access port

Sensor cable with key
positioned up away from pipe

Advertising