10.1.5 Communications

Menu

Ethernet

Option is used to view and set the IP address and Subnet Mask of the
transmitter.

Serial Options

Used for setting up the ‘Baud Rate’ and the serial communications to
‘RS232’ or ‘RS485’.

HART

Settings are used for configuring the protocol when using HART.
Default is for single device addressing.

Reset Comms

Used to re-initialize the communications ports without loosing data
history (as would happed by re-initializing the transmitter). This can
be used to restart acquiring data if, for example, the ethernet cable
was removed from the ethernet port while the StBasic program was
still running.

10.1.6 Diagnostics

Menu

Sensor Check

Selecting ‘Diagnostics>Sensor Check’ on the transmitter menu will
perform tests on all ‘ON’ sensors. (Individual sensors may be turned
off in the ‘Customize>Sensor Setup>State’ menu.) The test will take
several seconds to perform. After the test is completed, the screen
will display results.

The following are examples of test results.

Test Results: PASS
1:OK

2:OK

3:OK

4:OK

5:OK

6:OK

7:OK

8:OK

Figure 61

Test Passes Display Example

In the above figure the results indicate all sensors pass the test.

Test Results: FAIL: Continuity
1:SWITCH WT/BK

2:SWITCH WT/BK

3:OK

4:OK

5:OK

6:OK

7:DISCONNECTED 8:OK

Figure 62

Test Failed