CiDRA GVF-100 Gas Volume Fraction Monitoring System User Manual

Page 92

Advertising
background image

Copyright © 2006 CiDRA Corporation

Page 11-3

20639-01 Rev 03

values for water at various temperatures. Once the new value has
been entered on Line 4, press ‘ENTER’.

If no change is made (or after a change in Specific Gravity has
been made) pressing the

key will next display ‘→SOS’ on Line 3

and the current value on Line 4. To change the Line 4 value press
ENTER’ and use the arrow keys to enter the new value. Water at
25 ºC (4910.4 ft/s) is the default. Appendix E lists values for water
at various temperatures. Once the new value has been entered on
Line 4, press ‘ENTER’.

Once the Fluid Properties have been entered press the ‘BACK
key and

key and ‘→Pressure’ will be displayed on Line 2 and the

current assumed process pressure set value on Line 4. Note: the
units are PSIg. To change the Line 4 value press ‘ENTER’ and
use the arrow keys to enter the new value. This will result in a
fixed pressure being used for GVF calculations. If a Pressure
Transducer will be used to input process pressure into the
transmitter, it is not necessary to enter a pressure.

Pressing the

key will next display ‘→Temperature’ on Line 2 and

the current assumed process temperature set value on Line 4.
Note: the units are user selectable to degrees C or F. To change
the Line 4 value press ‘ENTER’ and use the arrow keys to enter
the new value. This will result in a fixed temperature being used
for GVF calculations. If a Temperature Transducer will be used to
input process temperature into the transmitter, it is not necessary
to enter a pressure.

Pressing the

key will next display ‘→Pressure Sel’ on Line 2. If

an external pressure sensor will not be used to input pressure to
the transmitter ‘Fixed’ appears on line 4 and use the

key to scroll

to the next menu item. However, if an external pressure sensor
will be used to input pressure to the transmitter press ‘ENTER’ and
use the

key to select either ‘Sensor #1’ or ‘Sensor #2’. (‘Sensor

#1’ or ‘Sensor #2’ refers to the sensor input the temperature
transducer is wired to on the terminal block within the transmitter
box.) Once the selection is made, press ‘ENTER’.

Pressing the

key will next display ‘→Temperature Sel’ on Line 2.

If an external temperature sensor will not be used to input
temperature to the transmitter ‘Fixed’ appears on line 4 and use
the

key to scroll to the next menu item. However, if an external

temperature sensor will be used to input pressure to the transmitter
press ‘ENTER’ and use the

key to select either ‘Sensor #1’ or

Sensor #2’. (‘Sensor #1’ or ‘Sensor #2’ refers to the sensor input
the temperature transducer is wired to on the terminal block within
the transmitter box.) Once the selection is made, press ‘ENTER’.

Pressing the

key will next display ‘→Altitude’ on Line 2 and the

selected value above (or below) sea level on Line 4. To change