Specifications Appendix A
Temperature Drift
The graphs below show the module’s temperature drift without autocalibration
for each thermocouple type over the thermocouple’s temperature range, assuming
terminal block temperature is stable. The effects of CJC temperature drift are
not included.
Figure 41 - Module Temperature Drift Using Type B Thermocouple
Figure 42 - Module Temperature Drift Using Type C Thermocouple
0
0.02
0.04
0.06
0.08
0.10
0.12
0
200
400
600
800
1000
1200
1400
1600
1800
2000
Thermocouple Temperature °C
Te
mperature Drift °C/°C
0
0.01
0.02
0.03
0.04
0.05
0.06
0.07
0.08
0.09
0.10
0
200
400
600
800
1000
1200
1400
1600
1800
2000
2200
2400
Thermocouple Temperature °C
Te
mperature Drift °C/°C