4 tubing, Tubing, Figure 7 – Metrohm 889 IC Sample Center User Manual

Page 24: Connecting pump and column

Advertising
background image

3.4 Tubing

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

889 IC Sample Center

1

2

3

4

5

6

4

2

5

3

1

Figure 7

Connecting pump and column

1

Connection to the syringe module
Port 3 of the injection valve.

2

Connection to the needle
Port 4 of the injection valve.

3

Connection to the high pressure pump
Port 1 of the injection valve.

4

Connection to the separation column
Port 6 of the injection valve.

5

Sample loop
Port 2 and 5 of the injection valve.

3.4

Tubing

The 889 IC Sample Center is equipped with the following standard tubing:

Tubing/Capillary

Material/Dimensions

Standard sample needle and
capillary (15 µL)

SS (inert-coated): 97 mm x 0.8 mm
OD x 0.25 mm ID

ETFE (Tefzel): 200 mm x 1/16" OD x
0.25 mm ID

Buffer tubing from the high
pressure valve to the syringe
valve (1000 µL)

ETFE (Tefzel): 1275 mm x 1/16" OD x
1.0 mm ID

Tubing from the syringe valve
to the washing bottle

PTFE: 400 mm x 1/18" OD x 1.6 mm
ID

Tubing from the syringe valve
to the outlet tubing

PTFE: 400 mm x 1/8" OD x 1.6 mm ID

Observe the following when mounting new tubing or capillaries:

Do not tighten the connectors excessively. This could block the flow
path.

Make sure that you always use tubing volumes which are suitable for
the other components in the flow path.