4 Functioning
889 IC Sample Center
■■■■■■■■
21
4.3
Full loop injection
The sample loop is filled completely (quantitatively) with sample. This type
of injection results in excellent reproducibility.
1 Initial situation
Figure 13
Full loop injection
The injection valve is in INJECT position. The sample needle pene-
trates the vial with the air needle. An overpressure channeled in
through the air needle ensures that no air or steam bubbles form
during the aspiration of the sample.
2 Rinsing the aspirating line