3 syringe unit, Syringe unit – Metrohm 889 IC Sample Center User Manual

Page 48

Advertising
background image

6.1 Error list

■■■■■■■■■■■■■■■■■■■■■■

40

■■■■■■■■

889 IC Sample Center

Problem

Cause

Remedy
If the error occurs again, please call the
Metrohm Service.

Error 313

The Home sensor for the
vertical movement of the
needle has not been acti-
vated.

1. Check whether the vertical movement of

the needle unit is obstructed.

2. If possible, remove possible obstacles.

If the error occurs again, please call the
Metrohm Service.

Error 315

The Home sensor for the
vertical movement of the
needle has not been deac-
tivated.

1. Check whether the vertical movement of

the needle unit is obstructed.

2. If possible, remove possible obstacles.

If the error occurs again, please call the
Metrohm Service.

Error 316

The Home sensor for the
vertical movement of the
needle has been unexpect-
edly activated.

Call the Metrohm Service.

Error 317

When moving the needle
unit vertically the needle
deflector has not recog-
nized a rack, nor a wash or
waste position.

If the error has occurred during moving to a
rack position: ensure that there is a rack with
vials or a micro titer plate on the rack holder.

If the error occurs again, please call the
Metrohm Service.

Error 318

When moving the needle
unit vertically the needle
deflector is blocked.

1. Check whether the movement of the nee-

dle deflector is obstructed.

2. If possible, remove possible obstacles.

If the error occurs again, please call the
Metrohm Service.

Error 319

The sample needle arm is
located in an undefined
vertical position.

Call the Metrohm Service.

6.1.3

Syringe unit

Problem

Cause

Remedy

Error 324

The syringe valve has not
reached the required port.

Call the Metrohm Service.

Error 330

The Home sensor of the
syringe has not been acti-
vated.

1. Check whether the flow path is kinked or

blocked in on one or more places.

2. If possible, free the flow path of blockages.