5.3 Maintenance and service

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

889 IC Sample Center

Sample rack. If sample has been spilled on the sample rack, clean this
with a soft cloth and a mild cleaning fluid.

Drainage lines. Rinse the drainage tubing regularly in order to pre-
vent blockage and to ensure the drainage of liquids and condensation
water.

5.3

Maintenance and service

5.3.1

Sample loop

The 889 IC Sample Center is equipped as standard with a 100 µL sample
loop. Other sample loops can be mounted. Please note however that the
correct combination of syringe and buffer loop is required in order to ach-
ieve good results.

Observe the following when mounting a sample loop:

Mount the sample loop at Port 2 and Port 5 of the injection valve.

Adjust the configuration settings in the control software for the
changed sample loop volume.

Note

The maximum injection volume is calculated in accordance with the fol-
lowing formulas:

Full loop: Injection volume = Loop volume

Partial loopfill: max. injection volume = 0.5 × loop volume

µL Pickup: max. injection volume = (loop volume – 3 × needle volume)
/ 2