8.3 Analytical characteristics

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

889 IC Sample Center

8.3

Analytical characteristics

Injection modes

Full-Loop, Partial Loopfill, Pickup mode, PASA

™ (pressure-assisted

sample aspiration)

Reproducibility

Relative standard deviation

≤ 0.3% for the Full-Loop injection.

Relative standard deviation

≤ 0.5% for the Partial Loopfill injection,

injection volume > 10 µL.
Relative standard deviation

≤ 1.0% for the Pickup injection, injection

volume > 10 µL.

Memory effect

< 0.05 % with programmable needle rinsing

8.4

Programming

Injection modes

Full-Loop, Partial Loopfill, Pickup Mode

Injection volume

1…5000 µL (with 1 µL increment), depending on the system settings.

Max. injection vol-
ume

Full Loop = Sample loop volume

Partial Loopfill = 0.5 x Sample loop volume

Pickup = (Sample loop volume - 3 x Needle volume)/2

8.5

Interfaces

Output signal

1 programmable relay outlet as injection mark

Data interface

USB connector, type B (for the connection to a PC)

8.6

Options (pre-installed)

Cooling for the
sample rack

Built-in Peltier element
Range: 4 °C to 3 °C below ambient temperature
Air temperature in the sample vessel: 4 °C

±2 °C (at the temperature

sensor)
(Temperature at 80% relative humidity and ambient temperature of
25°C)