3.6 Rinsing the syringe

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

889 IC Sample Center

Condensation and leakage outlet
All leakage liquid and condensation water (from the cooling module) will
be channeled through the right-hand tubing connector. If the cooling
option is used, it is advisable to connect this connector with a waste con-
tainer underneath the 889 IC Sample Center.

Figure 11

Mounting the leakage tubing

3.6

Rinsing the syringe

The syringe must be rinsed bubble-free prior to the start-up of the 889 IC
Sample Center.

Fill the washing bottle with isopropanol.

Warning

Do not use any saline or buffer solutions. Crystals could block or dam-
age the entire system.

Proceed as follows:

1 Immerse the end of the tubing for the washing bottle into the vessel.

2 Open the manual control in MagIC Net

™.

3 Select 889 IC Sample Center in the device selection (all devices).

4 On the tab General, carry out the function Washing.

The syringe, the needle and the tubing are rinsed thoroughly.

5 If needed, carry out the function a second time until the syringe no

longer contains any bubbles and the tubing is filled.

6 Repeat the washing procedure with ultra pure water or possibly with

eluent.