7 mains connection, 8 safety specifications, 9 electromagnetic compatibility (emc) – Metrohm 889 IC Sample Center User Manual

Page 57: Mains connection, Safety specifications, Electromagnetic compatibility (emc)

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

8 Technical specifications

889 IC Sample Center

■■■■■■■■

49

8.7

Mains connection

Voltage

100…240 V ±10 %

Frequency

50 / 60 Hz

Power consump-
tion

200 VA

Fuse

2.5 ATH

8.8

Safety specifications

Design and testing

According to EN/IEC 61010-1, UL 61010-1, CAN/CSA-C22.2 No.
61010-1, protection class

I.

Safety instructions

This document contains safety instructions which have to be followed
by the user in order to ensure safe operation of the instrument.

8.9

Electromagnetic compatibility (EMC)

Emission

Standards fulfilled:

EN/IEC 61326-1

EN/IEC 61000-6-3

FCC 47 CFR Part 15

EN/IEC 61000-3-2

EN/IEC 61000-3-3

Immunity

Standards fulfilled:

EN/IEC 61326-1

EN/IEC 61000-6-2

EN/IEC 61000-4-2

EN/IEC 61000-4-3

EN/IEC 61000-4-4

EN/IEC 61000-4-5

EN/IEC 61000-4-6

EN/IEC 61000-4-8

EN/IEC 61000-4-11