2 needle unit, Needle unit – Metrohm 889 IC Sample Center User Manual

Page 47

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Troubleshooting

889 IC Sample Center

■■■■■■■■

39

Problem

Cause

Remedy

Error 297

The Home sensor of the
rack holder has unexpect-
edly been activated.

Call the Metrohm Service.

Error 298

The rack is located in an
undefined position.

Carry out the function [Reset device] in the
Manual Control of the sample changer.

If the error occurs again, please call the
Metrohm Service.

6.1.2

Needle unit

Problem

Cause

Remedy

Error 303

The needle is located in an
undefined horizontal posi-
tion.

Carry out the function [Reset device] in the
Manual Control of the sample changer.

If the error occurs again, please call the
Metrohm Service.

Error 304

The Home sensor for the
horizontal movement of
the needle has not been
activated.

1. Check whether the horizontal movement

of the needle is obstructed.

2. If possible, remove possible obstacles.

If the error occurs again, please call the
Metrohm Service.

Error 306

The Home sensor for the
horizontal movement of
the needle has not been
deactivated.

1. Check whether the horizontal movement

of the needle is obstructed.

2. If possible, remove possible obstacles.

If the error occurs again, please call the
Metrohm Service.

Error 307

The Home sensor for the
horizontal movement of
the needle has unexpect-
edly been activated.

Call the Metrohm Service.

Error 308

In order to reach the hori-
zontal Home position of
the needle, the expected
number of steps has not
been carried out.

1. Check whether the horizontal movement

of the needle is obstructed.

2. If possible, remove possible obstacles.

If the error occurs again, please call the
Metrohm Service.

Error 312

The needle is located in an
undefined vertical position.

Carry out the function [Reset device] in the
Manual Control of the sample changer.