Metrohm 850 Professional IC Anion MSM-HC - LP-Gradient User Manual

Page 23

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

850 Professional IC Anion MSM-HC LP Gradient

■■■■■■■■

13

Place the detector in the instrument and connect it (see manual
of the detector)
.

Remove transport locking screws (see "Removing transport lock-
ing screws", page 23)
.

Connect the leak sensor (see "Connecting the leak sensor ", page
23)
.

Connect drainage tubings (see "Installing drainage tubings",
page 26)
.

3 Installing the eluent path

Equip the eluent aspiration tubing and connect with the eluent
bottle (see Chapter 3.9.1, page 29).

Instead of the column, connect the 6.2744.040 coupling with a
6.2744.070 short pressure screw at the end of the pre-installed
column inlet capillary.

Use a 6.2744.070 pressure screw to connect the capillary of the
MSM-HC labeled with in to the other end of the 6.2744.040 cou-
pling.

Connect the detector inlet capillary and the capillary of the
MSM-HC labeled with out to one another using one 6.2744.040
coupling and two 6.2744.070 short pressure screws (see manual
of the detector
).

4 Installing the sample path

Note

The sample degasser does not have to be connected. We recom-
mend the usage of the sample degasser only if the sample matrix
requires it.

Guide the 6.1803.040 sample aspiration capillary connected to
the sample input of the injection valve out of the instrument
through a capillary feed-through and connect it with the Sample
Processor or

If the sample degasser is used connect the sample aspiration
capillary with a 6.2744.090 long pressure screw to the input of
the sample degasser (see Chapter 3.15, page 45).
Connect a part of the 6.1803.040 sample aspirating capillary with
a 6.2744.090 long pressure screw to the input of the sample
degasser. Guide the other end out of the instrument through a
capillary feed-through and connect it to the Sample Processor.

Advertising