3 Installation
850 Professional IC Anion MSM-HC LP Gradient
■■■■■■■■
41
1
2
3
Figure 21
Mixing coil for low pressure gradients connection
1
Inlet valve holder
2
PEEK pressure screw, short 6.2744.070
3
Mixing coil for low pressure gradients
6.2758.020
3.12.3
Deaerating the high pressure pump
The high pressure pump will only operate perfectly if the pump head con-
tains no more air bubbles. Therefore it must be deaerated during initial
start-up and after every change of eluent.
Caution
The high pressure pump must not be deaerated before the initial start-
up (see Chapter 4.1, page 64).
Deaerate the high pressure pump as follows (see Figure 22, page 42):
Deaerating the high pressure pump
The instrument must be connected to the PC and switched on to deaerate
the high pressure pump.