3.8 Capillary and cable feed-throughs

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

850 Professional IC Anion MSM-HC LP Gradient

7

Drainage tubing
Section of the 6.1816.020 silicon tubing.
Guides escaped liquid into a waste con-
tainer.

8

Drainage tubing connector
For draining the escaped liquid from the
base tray through the connected drainage
tubing.

9

Drainage tubing connector
For supplying escaped liquid through the
connected drainage tubing to the leak sen-
sor.

Proceed as follows to install the drainage tubings:

Installing drainage tubings

1 Plug drainage tubing (8-2) into the drainage tubing connector (8-1)

of the flask holder and shorten to required length.

2 Plug drainage tubing (8-4) into the drainage tubing connector (8-3)

of the detector chamber and shorten to required length.

3 Connect drainage tubing (8-2) from the flask holder and drainage

tubing (8-4) from the detector chamber to the Y connector (8-5).

4 Connect drainage tubing (8-6) to the Y connector (8-5), shorten to

required length and plug the other end of the drainage tubing into
the drainage tubing connector (8-9) of the base tray.

5 Plug drainage tubing (8-7) into the drainage tubing connector (8-8)

of the base tray and guide the other end into a waste container.

3.8

Capillary and cable feed-throughs

Several openings have been integrated for feeding through capillaries and
cables. They are located on the doors (see Figure 9, page 27), on the
rear panel or below the flask holder or above the base tray (see Figure 10,
Page 28)
.