3 Installation

850 Professional IC Anion MSM-HC LP Gradient

■■■■■■■■

45

3

4

5

2

3

6

1

2

Figure 24

Pulsation damper – Connection

1

Connection capillary
Connection to the inline filter.

2

Fastening screws

3

PEEK pressure screws, short
6.2744.070

4

Holder for pulsation damper

5

Pulsation damper 6.2620.150

6

Connection capillary
Connection to the injection valve.

3.15

Sample degasser

The sample degasser removes gas bubbles and disolved gases from the
sample. For degassing, the sample flows into a vacuum chamber through
a special fluoropolymer capillary.

Gas bubbles in the sample lead to poor reproducibility, as the quantity of
sample in the sample loop would not always be the same. Samples (con-
taining gas) should therefore be degassed before injection. For this the
sample is sucked through a degasser chamber before injection, whereby
any gas bubbles are automatically removed.

Note

When using the sample degasser, the rinsing time extends by at least 2
minutes.