3.2 Initial installation

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

850 Professional IC Anion MSM-HC LP Gradient

5 Installing the peristaltic pump

See Chapter 3.19.2, Page 56

Install pump tubings.

Connect aspirating capillaries for the regeneration solution and
the rinsing solution.

6 Installing the MSM-HC

Connect the MSM-HC capillary labeled with regenerant to the
outlet of the pump tubing for the regeneration solution.

Connect the MSM-HC capillary labeled with rinsing solution to
the outlet of the pump tubing for the rinsing solution.

Guide the two MSM-HC capillaries labeled with Waste reg. and
Waste rins., respectively, through an instrument's capillary feed-
through to the waste container and fix them there.

7 Connecting the instrument

Connect the instrument to the PC (see Chapter 3.20.1, page
60)
.

Connect instrument to mains supply (see Chapter 3.20.2, page
60)
.

8 Initial start-up

See Chapter 4.1, Page 64

Switch on the PC and start MagIC Net.

Switch on and prepare the instrument.

Deaerate the high pressure pump.

Rinse the instrument without column(s).

9 Connecting the columns

Remove the 6.2744.040 coupling between the column inlet capil-
lary and the MSM-HC capillary labeled with in.

Connect guard column (optional) (see Chapter 3.21, page 61)

– Fasten the guard column to the end of the column inlet

capillary according to the specifications contained in the
leaflet accompanying the guard column.

– Rinse guard column.