5 Operation and maintenance

850 Professional IC Anion MSM-HC LP Gradient

■■■■■■■■

71

Pump head maintenance
An unstable baseline (pulsation, flow fluctuations) is in many cases the
result of contaminated valves (40-2), (40-3) or defective, leaking piston
seals on the high pressure pump. Proceed as follows for cleaning contami-
nated valves and/or replacing worn parts such as pistons, piston seal and
valves:

This maintenance work should be carried out at least once a year.

Removing the pump head

1 Turn off high pressure pump and wait until pressure is released.

2 Loosen the pressure screw on the inlet valve holder (20-2) and

unscrew the connection tubing together with the coupling (20-9)
pump head input capillary (20-7) and eluent aspiration tubing from
the pump head.

3 Unscrew the pump head output capillary (20-13) from the pump

head.

4 Remove pump head from the pump housing by loosening the 4 fas-

tening screws (20-5) using the 6.2621.030 hexagon key. The main
piston is on the left (viewed from the front), and the auxiliary piston
is on the right.

Cleaning/replacing the zirconium oxide piston

Clean one piston after the other as follows:

1 Removing the piston cartridge from the pump head

Loosen the piston cartridge with a wrench and unscrew from the
pump head by hand.