Figure 34, Removing piston – Metrohm 850 Professional IC Anion MSM-HC - LP-Gradient User Manual
Page 82
Advertising
5.5 High pressure pump
■■■■■■■■■■■■■■■■■■■■■■
72
■■■■■■■■
850 Professional IC Anion MSM-HC LP Gradient
2
1
Figure 34
Removing piston
1
Pump head
2
Piston
2 Dismantling the piston
Caution
On the inside of the piston cartridge there is a taut spring than can
jump out of the piston cartridge if suddenly loosing tension.
When opening the piston cartridge, hold pressure towards the
spring and unscrew carefully.
■
Loosen the screw of the piston cartridge with a wrench and
unscrew carefully by hand and by holding pressure towards the
taut spring.
■
Remove the zirconium oxide piston and lay on a tissue.
■
Remove the spring retainer, spring and the inner plastic sleeve
from the piston cartridge and lay by.
■
Remove the backup ring from the pump head and lay to the other
parts.