Figure 16, Eluent degasser – Metrohm 850 Professional IC Anion MSM-HC - LP-Gradient User Manual

Page 44

Advertising
background image

3.10 Eluent degasser

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

850 Professional IC Anion MSM-HC LP Gradient

Connecting the eluent degasser

5

6

1

2

3

4

Figure 16

Eluent degasser

1

Eluent degasser input

2

Eluent degasser output

3

Tubing flare
With tubing nipple.

4

Clamping screw

5

Eluent aspiration tubing 6.1834.080
For aspirating the eluent. The clamping
screw (16-

4

) is firmly mounted.

6

Connection capillary 6.1834.090
Connection from the eluent degasser to the
high pressure pump (see Chapter 3.12, page
39)
. The clamping screw
(16-

4

) is firmly

mounted.

1

Caution

The clamping screws (16-4) must be tightened carefully. Use the
6.2621.050 wrench for this.

Insert the eluent aspiration tubing (16-5) in the eluent degasser
input (16-1).

Carefully tighten clamping screw (16-4).

2

Insert connection capillary (16-6) (the end with the longer clamp-
ing screw (16-4)) into the eluent degasser output (16-2).

Carefully tighten clamping screw (16-4).

Connect the other end of the connection capillary (16-6) (with
the shorter clamping screw ) to the high pressure pump .

Advertising