3 Installation

850 Professional IC Anion MSM-HC LP Gradient

■■■■■■■■

63

each column. You can request detailed information on special IC applica-
tions in the corresponding "Application Bulletins" or "Application
Notes
", available in the Internet at

http://www.metrohm.com

in the

Applications area or via the Metrohm agent responsible free of charge.

Note

The separation column may only be installed after the initial start-up
(see Chapter 4.1, page 64) of the instrument. Until then, use the
6.2744.040 coupling instead of the guard and separation column.

Connecting and rinsing the separation column

1 Connecting the separation column

Caution

When using the columns, always ensure that these are correctly
inserted corresponding to the flow direction indicated.

Remove stoppers from the separation column.

Connect the lower end of the separation column with a
6.2744.070 short PEEK pressure screw to the outlet capillary of
the guard column (if used), or to the column input capillary .

2 Rinsing the separation column

Place beaker under the outlet end of the separation column.

Set the flow rate of the high pressure pump to the value suitable
for the separation column selected.

Start the high pressure pump and rinse the separation column
approx. 10 minutes with eluent.

Switch off the high pressure pump again.

3 Mounting the separation column

Hang separation column with chip into the column holder.

Note

The iColumns are equipped with a chip on which their operating data is
saved. The chip has to be hooked into the chip holder provided for this
so that the column recognition can function.