4.7 Batch reprocessing

790 Personal IC

149

4.7.3

Batch reprocessing queue editor

Open batch reprocessing queue editor window

REPROCESS /

<Edit sample table>

This menu item opens the batch reprocessing editor program for edit-

ing the batch reprocessing table.

No

Row

number.

File name

Name of the chromatogram file (read-only).

Method

Name of the method file

*.mtw

used for re-

cording the chromatogram (read-only).

Ident

User defined identifier for chromatogram. Will

be placed into appropriate passport field when

starting the chromatogram.

Vial

Autosampler vial position to take sample from.

Will be placed into appropriate passport field.

Volume

Injected volume of sample in µL. Will be placed

into appropriate passport field.

Dilution

Sample dilution prior to injection. Will be placed

into appropriate passport field.

Amount

Sample amount. If this value is different for the

calibration run (c) and the sample run (s), the

component concentrations of the sample are

calculated as follows:

C

s

=

C

c

Amount

s

/

Amount

c

Will be placed into appropriate passport field.

Internal standard amount

Concentration of the internal standard compo-

nent for relative concentration calculations. Will

be placed into appropriate passport field.

Calibration level

Calibration level (see section 4.4.5) for the sam-

ple: Level

0

stays for normal analysis, levels

1

and greater for calibration runs. Correct filling of

this column enables to calculate calibration

coefficients automatically.