5.2 Maintenance and servicing

790 Personal IC

155

5.2

Maintenance and servicing

5.2.1 General

information

Care

The 790 Personal IC requires proper care and attention. Excessive

contamination of the instrument could possibly lead to malfunctions

and a shorter service life of the inherently rugged mechanical and

electronic parts.
Spilled chemicals and solvents should be wiped up immediately. It is

especially important to protect the plug connections at the rear of the

instrument (particular the mains plug) against contamination.

Although constructional measures have been designed to virtually

eliminate such a situation, should corrosive media penetrate the

interior of the instruments the mains plug of the 790 Personal IC must

be immediately disconnected to prevent extensive damage to the

instrument electronics. Inform Metrohm service if your instrument(s)

have been damaged in such a way.

The instrument must not be opened by untrained personnel. Please

comply with the safety notes in section 1.4.1.

Maintenance by Metrohm service

Maintenance of the 790 Personal IC is best done as part of an annual

service performed by specialists from the Metrohm company. If work is

frequently performed with caustic and corrosive chemicals, it may be

necessary to shorten the interval between servicing.
The Metrohm service department is always willing to offer expert advice

on the maintenance and servicing of all Metrohm instruments.

5.2.2 Passivation

Passivation of the complete IC System (without column) by rinsing with

20...50 mL 0.2 mol/L HNO

3

is only required when unusual alterations to

the measuring properties of the cell are observed. For passivation, the

separating column 63

63

63

63 is removed from the 790 Personal IC. The two

capillaries 20

20

20

20 and 37

37

37

37 (see Fig. 13 and Fig. 15) are directly connected to

each other using the Coupling 26

26

26

26 (6.2620.060).