2.8 Suppressor module

790 Personal IC

41

4 Connect aspirating tubing for H

2

O

• Prepare aspirating tubing 71

71

71

71: Cut a piece of the 6.1803.020

PTFE tubing to the required length (normally ca. 120 cm).

• Insert one end of the aspirating tubing 71

71

71

71 into the rinsing

solution storage bottle and tighten it so that the tubing is

firmly held.

• Insert the free end of aspirating tubing 71

71

71

71 into one of the

openings 5

555 of the 790 Personal IC (see Fig. 2) and pull it suf-

ficiently far into the interior.

• Mount a coupling 73

73

73

73 (6.2744.030) to the inlet end of the front

pump tubing 75

75

75

75.

• Mount a 6.2744.010 Compression fitting at the end of the

aspirating tubing 71

71

71

71 and screw this compression fitting on to

the coupling 73

73

73

73 (see Fig. 15).

2.8.4

Connection of the suppressor module

The three inlets and outlets numbered 1...3 on the suppressor module

39

39

39

39 each have 2 permanently mounted PTFE capillaries, which must be

connected as follows (see Fig. 15 and Fig. 16).

1 Inlet capillary for eluent

• Screw inlet capillary 78

78

78

78 marked with "Eluent" of suppressor

module 39

39

39

39 to outlet end of separating column 63

63

63

63 using a

6.2744.010 Compression fitting.

2 Outlet capillary for eluent

• Screw outlet capillary 79

79

79

79 marked with "Detector" of suppres-

sor module 39

39

39

39 to coupling 26

26

26

26 using a 6.2744.010 Compres-

sion fitting.

• Screw inlet capillary 37

37

37

37 of detector block 38

38

38

38 to other end of

coupling 26

26

26

26.

3 Inlet capillary for H

2

SO

4

• Attach inlet capillary 81

81

81

81 marked with "H

2

SO

4

" of suppressor

module 39

39

39

39 using a 6.2744.010 Compression fitting to the filter

unit PEEK 28

28

28

28 connected to the rear pump tubing 74

74

74

74.

4 Outlet capillary for H

2

SO

4

• Pull outlet capillary 83

83

83

83 marked with "Waste" of the suppressor

module 39

39

39

39 from below through one of the openings 6666 out of

the inner compartment of the 790 Personal IC.

• Lead outlet capillary 83

83

83

83 to a sufficiently large waste container

and fix it in place.