3 connecting the eluent bottle – Metrohm 790 Personal IC User Manual

Page 36

Advertising
background image

2.6 High-pressure pump

790 Personal IC

27

2.6.3

Connecting the eluent bottle

To connect the eluent bottle with the high-pressure pump, insert the

supplied 6.1834.010 aspirating tubing (i.d. = 1.5 mm, e.d. = 2.5 mm,

length = 2.5 m) into one of the openings 5

555 or 7777 in the interior of the 790

Personal IC. Pull the aspirating tubing sufficiently far into the interior of

the 790 Personal IC and push at least 5 mm of it onto aspirating capil-

lary 32

32

32

32 (see Fig. 3 and Fig. 4) of the high-pressure pump (it may be

necessary to use emery paper). Screw on the 6.2821.090 aspirating fil-

ter at the other end of the aspirating tubing and insert it into the eluent

bottle.

Only degassed (with N

2

, He or vacuum) and microfiltered (0.45

µ

m

filter) eluents should be used!
Care must be taken that the eluent used is freely miscible with any

solvent remaining in the pump head (the pump head leaves the

factory filled with either isopropanol or methanol/water). If this is not

the case then the pump must first be rinsed with a solvent which is

miscible with both the previous eluent and the following eluent (e.g.

acetone).





2.6.4

Deaerating the pump and rinsing the pulsation dampener

The first time that it is started up the high-pressure pump must be

deaerated. Proceed as follows:

1 Prepare for deaeration

• Open the rotary knob on purge valve 31

31

31

31 by approx. ½ turn in

the counterclockwise direction (see Fig. 3 and Fig. 4).

• Mount a PEEK compression fitting 46

46

46

46 (see section 2.3.4) to

the end of capillary 30

30

30

30 and screw coupling 24

24

24

24 (6.2744.120) to

the compression fitting 46

46

46

46.

• Push 6.2816.020 Syringe (without needle) into the connection

on coupling 24

24

24

24 until the stop is reached.

2 Open system

• Start the «790 Personal IC» PC program, if it has not been

already been started (see section 2.5.3).

• Select

File / Open / System

in the main window. Open either the

subfolder

790com1

(if the 790 Personal IC is connected to

COM1) or the subfolder

790com2

(if the 790 Personal IC is

connected to COM2). Select the system file

Anion.smt

or

Cation.smt

and click on

<Open>

. The corresponding system

window is opened: