5 Notes – Maintenance – Faults

790 Personal IC

170

5.4

Diagnostic tests / Validation / GLP

The requirements of

GLP (Good Laboratory Practice) include a peri-

odic check of analytical measuring instruments with regard to their re-

producibility and accuracy using

Standard Operating Procedures,

SOP. Under the title «Application Bulletin No. 277 – Validation of

Metrohm Ion Chromatography Systems by using Standard Op-

erating Procedures (SOP)» an example of such a standard operating

procedure is available from Metrohm; it can be adapted and used with

the 790 Personal IC.
Further information on the subjects of QA, GLP and validation can also

be found in the brochure «

Quality management with Metrohm»,

which is available from your local Metrohm agency.
Testing of the electronic and mechanical function groups of Metrohm

instruments can and should be performed as part of a regular service

by trained personnel of the manufacturing company (see section 5.2.1).

All Metrohm instruments are equipped with start-up-test routines that

check for perfect functioning of the relevant assemblies when the in-

strument is switched on. If no error message is displayed, it may be as-

sumed the instrument is operating without faults.
The Metrohm company also supplies its instruments with an integrated

diagnostic program which, in the case of possible malfunctions or faulty

behavior, allows the service technician to check the functioning of cer-

tain assemblies and localize the fault.