8 cleaning the suppressor – Metrohm 790 Personal IC User Manual

Page 172

Advertising
background image

5.2 Maintenance and servicing

790 Personal IC

163

pump.

• Set the flow at the high-pressure pump to 0.5 mL/min and

rinse the suppressor unit with 1 mol/L H

2

SO

4

for 5 to 10 min.

• As the pressure falls, slowly increase the flow at the high-

pressure pump to 2 mL/min. Do not exceed a maximum

pressure of 2 MPa (20 bar).

• Switch off high-pressure pump.

• Switch suppressor to the next position using the

button.

• Connect the inlet capillary 80

80

80

80 marked "H

2

O" to the inlet capil-

lary 18

18

18

18 using coupling 26

26

26

26 (see Fig. 10 and Fig. 15).

• Set the flow at the high-pressure pump to 0.5 mL/min and

rinse the suppressor unit with 1 mol/L H

2

SO

4

for 5 to 10 min.

• As the pressure falls, slowly increase the flow at the high-

pressure pump to 2 mL/min. Do not exceed a maximum

pressure of 2 MPa (20 bar).

• Switch off high-pressure pump.

3 Connect suppressor to IC system

• Connect inlet capillaries 80

80

80

80 and 81

81

81

81 to the peristaltic pump

(see section 2.8.4).

• If the pressure problems persist, replace the suppressor rotor

(see section 5.2.9).



5.2.8

Cleaning the suppressor

It may be necessary to clean the suppressor in the following cases:
• High counter pressure on the suppressor connection

tubing

• Irremediable blockage of the suppressor (the suppres-

sor can no longer deliver solutions)

• Irremediable obstruction of the suppressor (the sup-

pressor can no longer be switched to next position)

To clean the connection piece and the rotor, proceed as follows (see

Fig. 20):

1 Disconnect suppressor from IC system

• Disconnect input capillary 78

78

78

78 of the suppressor module 39

39

39

39

from the separating column 63

63

63

63 (see Fig. 15).

• Disconnect output capillary 79

79

79

79 from inlet capillary 37

37

37

37.

• Disconnect inlet capillaries 80

80

80

80 and 81

81

81

81 from the filter units 28

28

28

28

(supply from peristaltic pump).