Metrohm 790 Personal IC User Manual

Page 45

Advertising
background image

2 Installation

790 Personal IC

36

70

70

70

70

19

19

19

19

19

19

19

19

37

37

37

37

38

38

38

38

22

22

22

22

63

63

63

63

70

70

70

70

20

20

20

20

Fig. 13: Installation of column without suppressor

2.7.8

Installation of the separating column with suppressor

With the 2.790.0020 Personal IC with suppressor module, the IC sepa-

rating column is first connected to the injector or precolumn. The con-

nection to the suppressor module and the detector block is described

in section 2.8.

1 Connect column to injector

• Remove end caps from column 63

63

63

63.

• Screw inlet end of separating column 63

63

63

63 (note flow direction)

to column connection capillary 20

20

20

20 or to the already installed

precolumn (procedure see section 2.7.7).

2 Rinse column

• Place a beaker beneath the column outlet.
• Open software window for manual system control.
• If necessary, modify

Flow rate

to the value suited for the

inserted separating column and click on

<Send to unit>

to

send this value to the 790 Personal IC.

• Switch on high-pressure pump (

IC pump

) by clicking

<On>

and

rinse column with eluent for ca. 10 min.

• Switch off high-pressure pump by clicking

<Off>

.

3 Fix column

• Insert one or two column holders 70

70

70

70 (6.2027.030, 6.2027.040

or 6.2027.050) in the mounting rails 19

19

19

19 and fasten separating

column 63

63

63

63 in the column holder 70

70

70

70.

19

19

19

19 Mounting rail

for column holder 70

70

70

70

20

20

20

20 Column connection

capillary

Connection injection

valve 22

22

22

22

separating column 63

63

63

63

22

22

22

22

Injection valve

37

37

37

37

Inlet capillary for

detector block

PEEK capillary, fixed

mounting

38

38

38

38 Detector block

(1.733.0110)

63

63

63

63 Separating column

70

70

70

70 Column holder

(6.2027.0X0)