2.4 Mains

connection

Follow the instructions below for connecting to the power supply. If

the instrument is operated with a mains voltage set wrongly and/or

wrong mains fuse, there is a danger of fire!



2.4.1

Setting the mains voltage

Before switching on the 790 Personal IC for the first time, check that the

mains voltage set on the instrument (see Fig. 9) matches the local

mains voltage. If this is not the case, you must reset the mains voltage

on the instrument as follows:

1 Disconnect mains cable

Disconnect mains cable from mains connection plug 12

12

12

12 of the

790 Personal IC.

2 Remove fuse holder

Using a screwdriver, loosen fuse holder 13

13

13

13 below the mains

connection plug 12

12

12

12 and take out completely.

3 Check and change fuse if necessary

Carefully take the fuse installed for the desired mains voltage out

of fuse holder 13

13

13

13 and check its specifications (the position of the

fuse in the fuse holder is marked by the white arrow imprinted

next to the mains voltage range):

100

120 V 1.0 A (slow-blow) Metrohm No. U.600.0016

220

240 V 0.5 A (slow-blow) Metrohm No. U.600.0013

4 Insert fuse

Change fuse if necessary and reinsert in fuse holder 13

13

13

13.

5 Install fuse holder

Depending on the desired mains voltage, insert fuse holder 13

13

13

13

in the 790 Personal IC so that the corresponding mains voltage

range can be read normally and the adjacent white arrow points

to the white bar imprinted below the fuse holder (see Fig. 9).