4.3 Systems

790 Personal IC

75

4.3.3 System

functions

Start/stop hardware and record baseline

SYSTEM / Control / Startup hardware (Measure Baseline)

Starting the hardware at the 790 Personal IC includes sending of

Sys-

tem startup values

, starting of the high-pressure pump and (if present)

starting of the peristaltic pump.
At the same time, the

recording of the measurement signal using

the method defined for the system is started. Independently of the set

chromatogram

Duration

, the measurement signal is recorded until a

new determination is started. Alternatively the baseline recording can

be stopped by clicking the icon of the chromatogram window. In

this case, the user is asked if the recorded baseline should be saved or

not.

SYSTEM / Control / Shutdown hardware

High-pressure pump and (if present) peristaltic pump at the 790 Per-

sonal IC are immediately stopped. A running determination and an ac-

tive sample queue are also stopped.

Start/stop determinations

SYSTEM / Control / Start determination

Start determination using the settings of the selected system. At this

start command, the

System startup values

are set at the 790 Personal IC.

The high-pressure pump and the peristaltic pump are started if they are

not already running. The time program is started immediately; the data

recording is started after switching the injection valve to the "Inject" po-

sition.

At each start of a determination, the

Edit sample description

window is

opened automatically. This window can be used for entering informa-

tion concerning the sample: