3 precolumn glass cartridges with cartridge holder – Metrohm 790 Personal IC User Manual

Page 41

Advertising
background image

2 Installation

790 Personal IC

32

2.7.3

Precolumn glass cartridges with cartridge holder

The precolumn glass cartridge METROSEP Anion Dual 1 (6.1006.030)

is inserted in the 6.2828.010 Precolumn cartridge holder as follows (see

Fig. 12):

1 Insert cartridge

• Remove end fittings 64

64

64

64 from screw caps 65

65

65

65.

• Insert precolumn cartridge 67

67

67

67 in sleeve 66

66

66

66 (mark flow direc-

tion of the column cartridge on sleeve, the column cartridge

should always be operated in the same flow direction).

• Screw both screw caps 65

65

65

65 loosely to sleeve 66

66

66

66 by hand.

• Screw both end fittings 64

64

64

64 into screw caps 65

65

65

65 so that their

capillary ends are seated in the PTFE seals of the cartridge.

• Tighten both screw caps 65

65

65

65 by hand.

• Tighten both end fittings 64

64

64

64 firmly using a screwdriver.

2 Connect precolumn

• Provide the column connection capillary 20

20

20

20 mounted on the

injection valve with a PEEK compression fitting 46

46

46

46 (see sec-

tion 2.3.4) and screw it tightly onto end fitting 64

64

64

64 on the inlet

side of the precolumn.

• Cut off as short a piece as possible from PEEK capillary 48

48

48

48

and fit this with two PEEK compression fittings 46

46

46

46 (see sec-

tion 2.3.4).

• Screw prepared capillary 48

48

48

48 to end fitting 64

64

64

64 on the other end

of the precolumn.

3 Rinse precolumn

• Place a beaker beneath the outlet capillary of the precolumn.
• Open software window for manual system control.
• If necessary, modify

Flow rate

to the value suited for the

inserted separating column and click on

<Send to unit>

to

send this value to the 790 Personal IC.

• Switch on high-pressure pump (

IC pump

) by clicking

<On>

and

rinse precolumn with eluent for ca. 10 min.

• Switch off high-pressure pump by clicking

<Off>

.


The rinsed precolumn can be connected directly to the separating

column instead of by means of a PEEK capillary. Proceed as follows:
Screw off PEEK capillary from end fitting 64

64

64

64 of the precolumn.

Screw off end fitting 64

64

64

64 from screw cap 65

65

65

65 of the precolumn.

Screw connection piece 68

68

68

68 onto screw cap 65

65

65

65.

Screw screw cap 69

69

69

69 of separating column 63

63

63

63 onto connection

piece 68

68

68

68.