2.2

Installation of the 797 VA Computrace Stand

797 VA Computrace / Hardware-Manual 8.797.8001EN

13

Any break in the grounding inside or outside the instrument can make
it a hazard!

Plug the mains cable into mains connection plug 16 of the 797 VA Computrace Stand
(see Fig. 2).

2.2.2

Switching the instrument on/off

The 797 VA Computrace Stand is switched on and off using mains switch 14. When
the instrument is switched on, the pilot lamp 1 lights up.

2.2.3

Connection to the PC

The 797 VA Computrace Stand is connected to the PC via USB Cable 6.2151.020.
Proceed as follows:

The 797 VA Computrace Stand must not be connected until the Soft-
ware is installed.

1 Software installation

• Switch on PC and start operating system (Windows™

2000, XP Professional

or Vista Professional)) without connection of the 797 VA Computrace via
USB cable.

• Insert installation CD into CD drive.
• If the autorun option for the CD drive is disabled, select

<Start>

and

Run

.

Browse for the

Setup.exe

file on the installation CD and click on

<OK>

.

• Click on "

797

" and follow the instructions given in the setup program. The

software package will be installed in the desired directory (the default direc-
tory is

Program Files/Metrohm/797 VA Computrace

).

2 Connection of the 797 VA Computrace

• Connect 797 VA Computrace to the PC using the 6.2151.020 USB cable

and switch on the 797 VA Computrace. The PC detects a new USB device
and starts the setup wizard. Insert installation CD into CD drive and follow
the wizard instructions always selecting the recommended default options.

• Start the 797 VA Computrace Software.

The setup wizard is started three times when installing the instrument
driver. All three installation steps must be conducted to ensure proper
operation.