2.2

Installation of the 797 VA Computrace Stand

797 VA Computrace / Hardware-Manual 8.797.8001EN

15

3 Insertion of 4-way microtip (option)

For automatic solution addition with Dosinos or Dosimats, the 6.1824.000 4-
way microtip has to be installed. Proceed as follows:

• Remove stopper from nipple 27 and insert 4-way microtip 30 into nipple as

far as it will go.

• Tighten nipple using a 6.2739.010 Wrench until the 4-way microtip can no

longer move.

• Pull the 4 lengths of PTFE tubing of the 4-way microtip in succession from

above through the opening 68 (connection of Dosinos or Dosimats see sec-
tion 2.8).

4 Install stirrer or RDE

in operation with MME:

• Insert stirrer (6.1204.200) in opening 63 as far as it will go.
• Stretch drive belt 32 (6.1244.020) between drive wheel 35 and drive shaft

of the stirrer.

in operation with RDE:

• Screw electrode tip 98 (6.1204.XXX) to drive shaft 99 (6.1204.210) (see

also section 2.4).

• Insert RDE in opening 63 as far as it will go.
• Stretch drive belt 32 (6.1244.020) between drive wheel 35 and drive shaft

99 of the RDE.

• Attach electrode cable 20 (WE) to the RDE: push cable lug under the screw

and then tighten screw firmly.

Fig. 6:

Measuring head arm

23 Measuring head arm

24 Measuring head

54 Opening

for auxiliary electrode 43 (6.0343.000 Pt
- auxiliary electr. or optional GC electr.
comprising 6.1241.020 Electrode holder
and 6.1247.000 GC tip)

59

60

61

62

65 66

67

68

69

70

64

63

54

24

57

58

23

55

56

Advertising