2.11

Control lines

797 VA Computrace / Hardware-Manual 8.797.8001EN

67

For software settings and measurement procedures: see Software Manual 797 sec-
tion 8.6
, and Online-Help of the <797 VA Computrace Software> section Brightener
Analysis with the 838 Advanced Sample Processor and LAT
.

2.11 Control

lines

Control lines for 863 Compact Autosampler, 766 Sample Processor, 838 Advanced
Sample Processor and 731 Relay Box are listed in the table below:

797 VA Computrace

863 Compact
Autosampler

838 Advanced

Sample

Processor

731 Relay Box

Pin 3 (End of sample)

Pin 10 (Input 3)

Pin 4 (End of sample table)

Pin 24 (Input 6)

Pin 5 (Start)

Pin 12 (Input 7)

Pin 6 (Set Control line)

Pin 22 (Input 2)

Pin 22 (Input 2)

Pin 7 (Siphoning)

Pin 7 (Output 9)

Pin 8 (Rinsing)

Pin 8 (Output 10)

Pin 17 (Scan Control line)

Pin 5 (Output 0)

2.12 Connection of peripherals

The USB Connections (19) USB1 and USB2 on the backside of the 797 VA Compu-
trace Stand serve as USB-distributor of the connected PC. Any USB-instrument (e.g.
printer or 846 Dosing Interface) can be operated with these connections.

For connection and installation of a particular instrument read the respective “Instruc-
tions for Use” of the instrument and operating system.