Metrohm 797 VA Computrace User Manual

Page 93

Advertising
background image

4.1

Technical data

797 VA Computrace / Hardware-Manual 8.797.8001EN

85

Dummy Cell

Use

Checking of the 797 VA Computrace
Determination of the signal-to-noise ratio

Connections

AE

auxiliary electrode

RE

reference electrode

WE-L

working electrode linear mode

(RC element)

WE-D

working electrode differential mode

(peak/wave)

Inert gas (in general nitrogen N

2

)

Use

Operation of MME de-aeration of sample solution (Not
needed for electroplating bath analysis)

Purity

min. 5.0 (99.999 %)

Required pressure

p = 1 ± 0.2 bar (this gas pressure results in a gas flow
rate of ca. 20 L/h)

Dosing devices (Option)

Type

800 Dosino (2.800.0010) + 807 Dosing unit
(6.3032.XXX),

700 Dosino (2.700.0020) + 807 Dosing unit
(6.3032.XXX),

685 Dosimat (2.685.0010) + 806 Exchange unit
(6.3026.XXX) + Cable (6.2134.030)

805 Dosimat (2.805.0010) + 806 Exchange unit
(6.3026.XXX)

connect via MSB 1…3 (or to a 846 Dosing Interface)

Number

1...3 (with a 846 Dosing Interface, 4 additional can be
connected)

Plug

Mini-DIN

Manual operation

Dispensing, filling, adjustment of feed and filling rate

MSB connections

(MSB = Metrohm Serial Bus)

Dosing devices

Connection of max. 3 Dosing devices (MSB1 to MSB3)
to 797 VA Computrace. With a 846 Dosing Interface,
four additional Dosing devices can be connected.

USB connections

USB Standard

USB1.1

Connection computer

With cable 6.2151.020

USB ports

2 USB downstream Ports (type A sockets), each
500 mA, for the connection of peripherals as printer,…

Remote connections

Remote Interface

D-Sub, 25 pin remote interface to control sample
changer and rinsing equipment.