2 Installation

797 VA Computrace / Hardware-Manual 8.797.8001EN

36

2.5 Reference

electrode

2.5.1 Construction

The complete reference electrode (RE) 26 comprises two parts:

6.0728.0X0 Ag/AgCl-reference system (100)

with ceramic diaphragm type D, diameter = 1 mm
6.0728.030 Reference system: LL-Ag/AgCl/c(KCl) =3 mol/L;
6.0728.020 Reference system: Ag/AgCl/c(KCl) = 3 mol/L;

supplied in a holder filled with c(KCl) = 3 mol/L

as

standard

6.0728.010 Reference system: Ag/AgCl

supplied dry (option)

6.1245.010 Electrolyte vessel (101)

with ceramic diaphragm type D, diameter=3 mm; holds a
second electrolyte solution (bridge electrolyte) and thus
forms with the 6.0728.020 Reference system a complete
reference electrode in the so-called double junction
construction.

The construction of the reference electrode and the designations of the individual
parts are shown in Fig. 13.

Fig. 13:

Construction of the reference
electrode

26 Reference electrode

100 Reference system (6.0728.0X0)

101 Electrolyte vessel (6.1245.010)

102 Electrical connection for cable

"RE"

103 Vent opening

104 Ag/AgCl filling

105 Electrolyte compartment with

internal electrolyte

106 Diaphragm support made of

PCTFE

107 Diaphragm

108 Vent opening

109 Electrolyte compartment with

bridge electrolyte

110 Diaphragm

100

26

104

107

102

103

105

106

101

108

109

110