Metrohm 797 VA Computrace User Manual

Page 36

Advertising
background image

2 Installation

797 VA Computrace / Hardware-Manual 8.797.8001EN

28

9 Connect Multi-Mode Electrode

• Screw FEP tubing 34 for the inert gas supply into connection 71 of the

Multi-Mode Electrode 21.

• Screw FEP tubing 42 for the inert gas supply into connection 72 of the

Multi-Mode Electrode.

• Attach electrode cable 20 (WE) to screw connection 88 of the Multi-Mode

Electrode push cable lug under the screw and then tighten screw firmly.

10 Pressurize the MME

• Switch on 797 VA Computrace Stand with mains switch 14 (the 797 VA

Computrace Stand must first be installed properly as described in section
2.2
).

• Start the VA Computrace program and click on

or

MAIN WINDOW / Utility

/ Computrace control

to open the

COMPUTRACE CONTROL

window. Then switch

on the inert gas supply to the 797 VA Computrace Stand by clicking on

DME

. This pressurizes the Multi-Mode Electrode 21 and the mercury begins

to flow slowly out of the capillary.

• Gently tap the MME with your finger (to remove any air bubbles) and allow

mercury to flow out of the capillary into the empty measuring vessel for
approx. 2 min.

• Fill measuring vessel 7 with 10 mL ultra pure water and add 1 drop KCl

solution (in pure water, mercury drops from the capillary only with difficul-
ty).

• Allow mercury to flow out of the capillary for ca. 2 min while checking the

drop formation: The drop time should be ca. 3 s.

11 Adjusting the sealing needle 75

• Turn slotted screw 74 using a suitable coin slowly in a clockwise direction

until the mercury flow stops.

• Open slotted screw slightly in an anticlockwise direction until the mercury

flow restarts.

• Gently tap the MME with your finger and turn the slotted screw very slowly

clockwise until the mercury flow just stops. (The tapping action is used to
knock off the mercury drop so that it is easier to see whether mercury con-
tinues to flow).

• Finally, turn slotted screw a quarter of a turn clockwise.

12 Checking the MME for leaks

• Switch on the dropping mercury electrode by selecting

DME

in the

COM-

PUTRACE CONTROL

window and clicking on

. The mercury drops

freely out of the capillary.

• Select

HMDE

and click on

. A single mercury drop is formed.

Knock this off by gently tapping the MME 21 with your finger and check
that the mercury flow has really stopped. Repeat this operation several
times.

• If mercury continues to flow, turn slotted screw 74 still further in a clock-

wise direction and repeat check.

• If it is not possible to stop the mercury flow, both the glass capillary 87 and

the sealing needle 75 have to be replaced (see section 2.3.9).

Advertising