2.8

Connection of Dosing devices

797 VA Computrace / Hardware-Manual 8.797.8001EN

41

700 Dosino and 800 Dosino

1 Mount Dosing unit on Dosino

• Procedure, see Instructions for Use 700, section 2.2, and Instructions for

Use 800, section 3.1 respectively.

2 Insert 4-way microtip in 797 VA Computrace Stand

(see section 2.2.4)

• Remove stopper from nipple 27 and insert 4-way microtip 30 into nipple as

far as it will go (see Fig. 3 and Fig. 6).

• Tighten nipple manually, so that the 4-way microtip can no longer move.
• Pull the 4 lengths of PTFE tubing of the 4-way microtip 26 in succession

from above through the opening 64.

3 Connect PTFE-Tubing to Dosing unit

• Screw connection nipple of the PTFE tubing of the

4-way microtip 30 on 1 (Port 1) of the Dosing unit
807.

4 Close unused PTFE tubing

• Screw a 6.1808.000 Coupling (accessory of 797 VA Computrace Stand) on

each unused PTFE tubing of the 4-way microtip 30.

• Screw a 6.1446.040 Dummy stopper (accessory of 797 VA Computrace

Stand) on each 6.1808.000 Coupling.

5 Initialize Dosino(s)

• Switch on the 797 VA Computrace Stand using the mains switch 14.
• Switch on the PC and start the 797 VA Computrace program.
• Open the

Dosinos

tab (if the Dosino is connected to the 846 Dosing Interfa-

ce: the

Dosing Interface

tab) of the window

General settings

and enter the

following configuration: