2 Installation

797 VA Computrace / Hardware-Manual 8.797.8001EN

44

used to dose the solution in the measuring vessel.

• Click <OK>.

• Click on

or

MAIN WINDOW / Utility / Dosino control

to open the

DOSINO

CONTROL

window.

• The connected Dosimat and the mounted Exchange Unit are automatically

identified.

• Click the

button to empty and refill the Exchange unit installed

on the Dosimat.

• Check if there are air bubbles left in the glass cylinder of the Exchange unit.

If this is the case, repeat the flushing procedure by clicking the

button.

• Close the

DOSINO CONTROL

window.


2.8.3

Change Dosing-/Exchange unit

Dosing unit 807

The Dosing unit mounted on the Dosino can be changed only in the exchange posi-
tion which is reached after filling. Please proceed as follows:

1 Fill Dosing unit

At the start of the VA Computrace program, the Dosing unit is automatically
filled. So this step is only necessary if the Dosino has already been used during
the running program session.

• Click on

or

MAIN WINDOW / Utility / Dosino control

to open the

DOSINO

CONTROL

window.

• Click the

button of the corresponding Dosino to fill the Dosing

unit installed on the Dosino.

2 Change Dosing unit

• Unscrew connection nipple of the PTFE tubing of the 4-way microtip 30

from connection 1 and take off old Dosing unit.

• Mount new Dosing unit on Dosino, the new Dosing unit is automatically

identified.

• Screw connection nipple of the PTFE tubing of the 4-way microtip 30 on 1

(Port 1) of the Dosing unit.

3 Initialize Dosino(s)

• Click on

or

MAIN WINDOW / Utility / Dosino control

to open the

DOSINO

CONTROL

window.

• Click the

button of the corresponding Dosino to empty and re-

fill the Dosing unit installed on the Dosino.

• Check if there are air bubbles left in the glass cylinder of the Dosing unit. If

this is the case, repeat the flushing procedure by clicking the

button.

• Close the

DOSINO CONTROL

window.

Advertising