Connection of 863 Compact Autosampler

797 VA Computrace / Hardware-Manual 8.797.8001EN

51


2.9.3 Software

settings

Before putting the 797 VA Computrace Stand with the 863 Compact Autosampler in-
to operation, the following settings have to be made in the "797 VA Computrace
Software
" software program:

1 Set dosing parameters

• Click on

797 VA COMPUTRACE

/ Settings / General settings

and select the

Dosi-

nos

tab (if the Dosino is connected to the 846 Dosing Interface: the

Dosing

Interface

tab) of the window

General settings

and enter the following con-

figuration:

• Click on

to display the currently connected Dosing devices.

• Set parameters of used Dosing devices.

• Define the addition or predose solutions for the desired method in the

DOSINOS

window (procedure see Software Manual, section 5.2).

• Fill the Dosing- or Exchange units of the Dosinos/Dosimats with the desired

solutions and make sure that there are no gas bubbles in the glass cylinders

using the

button in the

DOSINO CONTROL

window (see section

7.2, Software Manual).

Note: If you choose Port 3 for menu item

Prep / Empty via port

, you must

install an FEP Tubing Connection 6.1805.XXX from Port 3 to a waste con-
tainer.