4 communication diagram dt, 2 installation, Fig. 40: communication diagram for dt – Metrohm 797 VA Computrace User Manual

Page 80: Measurement

Advertising
background image

2 Installation

797 VA Computrace / Hardware-Manual 8.797.8001EN

72

2.13.4

Communication diagram DT

797 VA Computrace

Dosinos 1..3

838 Advanced Sample

Processor

731 Relay Box

Set line: Start 838 (TTL 3)

Ready (TTL 7)

Set line: Siphoning (TTL 5)

Prep Dosino 1,2 ?

Prep Dosino 1, 2

YES

Set line: Siphoning (TTL 5)

Wait siphoning time

No. of Prep
cycles

NO

Set line: Immerse needle into sample (TTL 4)

Ready (TTL 7)

Predose

Measurement

Set line: Siphoning (TTL 5)

Wait siphoning time

Set line: Move to rinsing vessel (TTL 4)

Ready (TTL 7)

Set line: Siphoning (TTL 5)

Wait siphoning time

Set line: Rinsing (TTL 6)

Wait rinsing time

No. of rinsing cycles

More samples ?

NO

YES

Set line: End of sample table (TTL 2)

DT Suppressors with calibration curve / DT Record calibration curve

<Use for predose> active ?

YES

NO

Scan line : 838 ready

(TTL 7)

Scan line : 838 ready

(TTL 7)

Scan line : 838 ready

(TTL 7)

Wait siphoning time

Prep Dosino 3 ?

Prep Dosino 3

YES

NO

No. of Prep
cycles

Set line: Siphoning (TTL 5)

Wait siphoning time

Set line: Rinsing (TTL 6)

Wait rinsing time

No. of rinsing cycles

Empty Dosino 3

Prep Dosino 3 ?

Prep Dosino 3

YES

NO

No. of Prep
cycles

Set line: Siphoning (TTL 5)

Wait siphoning time

Export/Print results

Set line: End of sample (TTL 1)

Note: Use <Use

for predose> to

dose VMS via

Dosino 1

Note: Is only

done, if Dosino 3

was preped before

Ti

m

el

in

e

End

Fig. 40:

Communication diagram for DT

Advertising