2.13

Communication diagrams for automation

797 VA Computrace / Hardware-Manual 8.797.8001EN

73

2.13.5

Communication diagram "RC Record response curve"

797 VA Computrace

Dosinos 1..3

838 Advanced Sample

Processor

731 Relay Box

Set line: Start 838 (TTL 3)

Ready (TTL 7)

Set line: Siphoning (TTL 5)

Prep Dosino 1,2,3 ?

Prep Dosino 1, 2, 3

YES

Set line: siphoning (TTL 5)

Wait siphoning time

Set line: Rinsing (TTL 6)

Wait rinsing time

No. of Prep
cycles

Is it the

first sample of the

sequence ?

NO

Set line: Siphoning (TTL 5)

Wait siphoning time

NO

YES

Set line: Sample transfer (TTL 4)

Ready (TTL 7)

Predose

Measurement

Set line: Siphoning (TTL 5)

Wait siphoning time

Set line: Rinse needle (TTL 4)

Ready (TTL 7)

Set line: Siphoning (TTL 5)

Wait siphoning time

Set line: Rinsing (TTL 6)

Wait rinsing time

No. of rinsing cycles

Export/Print results

Set line: End of sample (TTL 1)

More samples ?

NO

YES

Set line: End of sample table (TTL 2)

RC Record response curve

<Use for predose> active ?

YES

NO

Scan line : 838 ready

(TTL 7)

Scan line : 838 ready

(TTL 7)

Scan line : 838 ready

(TTL 7)

No. of rinsing cycles

Wait siphoning time

Set line: Siphoning (TTL 5)

Wait siphoning time

Note: Use <Use

for predose> to

dose Electrolyte

via Dosino 1

Ti

m

e

line

End

Fig. 41:

Communication diagram for "RC Record response curve"