4 operation of the 863 compact autosampler, Operation of the 863 compact autosampler – Metrohm 797 VA Computrace User Manual

Page 60

Advertising
background image

2 Installation

797 VA Computrace / Hardware-Manual 8.797.8001EN

52

2 Set automation parameters

• Select the

Automation

tab.

• Select „813/863 Compact Autosampler“ for the field Sample processor in

the Sample handling-part of the

Automation

tab.

• Check the checkbox

Relay box

.

• Normally, the default settings can be used for samples with 10 mL volume:

3 Test automation parameters

• Fill two sample vessels with water and place them in position 1 and 2 on

the sample rack of the 863 Compact Autosampler.

• Click on

797 VA COMPUTRACE

/ Settings / General settings

and select the

Auto-

mation

tab.

• Click on

, check the automation parameters and modify them if

needed.

2.9.4

Operation of the 863 Compact Autosampler

After installation of the instruments according to sections 2.9.1...2.9.3 sample series
using the 863 Compact Autosampler can be started. Proceed always in the following
sequence:

1 Switch on instruments

• Switch on PC.
• Switch on 797 VA Computrace Stand.
• Switch on 863 Compact Autosampler and 843 Pump Station.